Responsive Respiratory EMS Oxygen Regulator - Preset 50 PSI

Responsive Respiratory EMS Oxygen Regulator - Preset 50 PSI

Brand/Manufacturer: RESPONSIVE RESPIRATORY INC
Hpfy Stores Save
  • Non Returnable
  • FSA Approved

Responsive Respiratory EMS Oxygen Regulator - Preset 50 PSI is intended for the administration of oxygen to patients who need to increase oxygen levels to improve their breathing conditions. In addition, some models may be used in emergency situations. In an emergency situation, one should be sure to contact a physician or EMS immediately. Preset 50 PSI EMS Regulator, CGA 540 comes with DISS. It is made up of lightweight aluminum outer sleeve with all brass inner core and is rated for use up to 3000 PSI. The oxygen regulator can be used on composite cylinders up to 3000 PSIG and has high visibility gauge. It meets FDA, CGA, ASTM Standards, including ignition testing.

 

Item # Desc Pkg Price
120-1260 50 PSI EMS Regulator Each
$71.49
$50.00

Responsive EMS Oxygen Regulator Features

  • Responsive Respiratory EMS Oxygen Regulator - Preset 50 PSI Set Up:
    - Inspect the cylinder following the instructions of the gas supplier. Be sure to purge the cylinder valve seat of debris
    - Inspect the regulator for oil or grease. If visible. DO NOT use the regulator
    - Gently thread the large nut on the regulator onto the connection on the cylinder valve
    - Tighten the nut with a wrench until snug
    - Attach the oxygen supply tubing to the outlet connection of the regulator
  • Instructions for use:
    - Be sure the regulator is in the off position by rotating the nob counterclockwise to the 0 position
    - For safety be sure you are not directly in front of or behind the regulator when opening the cylinder valve
    - Slowly turn the cylinder valve on (counterclockwise), about one full turn
    - If hissing sound comes, there is a leak in the system. Turn the cylinder off and turn the regulate on to relieve any built up pressure. Try tightening the regulator to the cylinder and opening the cylinder again. If this does not help, shut the system down again and contact a service representative
    - Adjust the flow rate setting by turning the knob clockwise until the desired setting shows through the window
    - If the cylinder pressure falls below 300 PSI, one should exchange the cylinder for a full one
    - Turn off the system, close the cylinder valve, wait for oxygen to flow out of the regulator, and turn the regulator to the 0 position
    - NEVER try to remove the regulator unless the cylinder is off and pressure has been relieved
  • Flow Ranges: Pre-set 50 PSI

  • Duration Chart

Warranty

  • 2 Year Manufacturers Warranty: Covers internal components only. External gauges, barbs, washers, knobs and Tee handles are not covered.