Monaghan AeroChamber Aerosol Holding Chamber for Mechanical Ventilation

Monaghan AeroChamber Aerosol Holding Chamber for Mechanical Ventilation

Brand/Manufacturer: MONAGHAN MEDICAL
Hpfy Stores Save
  • HyGiene Products
  • Non Returnable
  • FSA Approved

The Monaghan AEROCHAMBER Aerosol Holding Chamber for Mechanical Ventilation (HC MV) is a tried-and-true method for administering aerosol medications to mechanically ventilated patients. This versatile product, which has been clinically tested in both adult and neonatal ventilator applications, can be used with a resuscitation bag or placed in the ventilator circuit, but must be removed from the circuit once the aerosol treatment has occurred.

 

Item # Desc Pkg Price
98550 Holding Chamber Each
$16.25
$11.95

Monaghan Valved Holding Chamber Features

  • Monaghan AeroChamber Aerosol Holding Chamber is proven standard for pMDI aerosol delivery for mechanically ventilated patients
  • Provides rapid delivery of aerosols that is vitally important for the treatment of the critically ill patient
  • See through chamber facilitates visualization of the aerosol spray
  • Comes with 15/22mm adapter that permits the use of this holding within a ventilator circuit or a resuscitation bag
  • Connector allows for easy attachment to either an endotracheal or tracheostomy tube
  • Latex-free
  • Single patient use
  • Easy to use
  • BPA-free

 

What to buy with Monaghan 98550 Holding Chamber

 

How to Use Monaghan Aerosol Holding Chamber

In Ventilator Circuit

  1. Attach the 15/22 mm adapterto the 15mm O.D. end of theAeroChamber HC MV.
  2. Disconnect the inspiratory limb at the Y-connector
  3. Attach the 22mm I.D. end of the AeroChamber HC MV to the Y-connector
  4. Attach the inspiratory limb to the 15/22 mm adapter of the AeroChamber HC MV
  5. Shake the MDI to properly mix the medication according to the manufacturer’s guide- lines.
  6. Insert the MDI canister into the actuator port of the AeroChamber HC MV
  7. After exhalation and just prior to the next inspiration, press down on the MDI to spray medication into the AeroChamber® HC MV. The medication will be carried into the patient’s lungs during the next inspiratory phase.
  8. For each puff prescribed by the physician, repeat steps 5 to 7.
  9. Following treatment, remove AeroChamber HC MV from circuit and reconnect inspiratory limb to Y-connector. Return the AeroChamber HC MV to it’s protective bag and store it properly near the patient’s bed.

With a Resuscitation Bag

  1. Attach the 15/22 mm adapter to the 15mm I.D. end of the AeroChamber HC MV
  2. Attach the output opening of the resuscitation bag to the 15 mm O.D. end of the AeroChamber HC MV.
  3. Shake the MDI canister to properly mix the medication according to the inhaler manufacturer’s guidelines
  4. Insert the MDI canister into the actuator port of the AeroChamber HC MV
  5. Disconnect the ventilator circuit from the patient’s endotracheal airway and at- tach the AeroChamber HC MV to the patient’s endotraecheal airway.
  6. Before squeezing the resuscitation bag, press down on the MDI to spray the medication into the AeroChamber HC MV. Immediately squeeze the resuscitation bag one or two times for each puff of medication that is needed and as presribed by the physician.
  7. For each puff prescribed by the physician, repeat steps 3 to 6.
  8. Following treatment, disconnect AeroChamber HC MV from patient and reconnect ventilator circuit to patient’s endotracheal airway. Return the AeroChamber® HC MV to it’s protective bag and store it properly near the patient’s bed.

 

Customer Reviews