Graham Field John Bunn Vacutec 800 EV2 Aspirator

Graham Field John Bunn Vacutec 800 EV2 Aspirator

  • Made in USA

Graham-Field Vacutec 800 EV2 Aspirator is designed for single patient, everyday use. Comes with easy-to-read gauge, easy-to-turn flow regulator knob and with a suction cup feet for stable operation. It comes with automatic shut -off and designed for most homecare suction applications.

Vacutec 800 EV2 Aspirator Highlights:

  • Low level noise at less than 55 dBA 
  • HEPA Bacteria Filter 
  • 60 watts power consumption

Warning WARNING: Cancer and Reproductive Harm - www.P65Warnings.ca.gov

"

 WARNING: Cancer and Reproductive Harm - www.P65Warnings.ca.gov
 

"

John Bunn Vacutec 800 EV2 Aspirator Features

 

How To Operate Graham Field 800 EV2 Aspirator?

  • Make sure the ON/OFF switch located on the side of the unit is in the OFF position. Plug the aspirator into a properly grounded power outlet. 
  • Switch the ON/OFF switch to I position (ON); the unit will start running. 
  • Adjust the vacuum level from 0 to 560 mmHg by turning the vacuum adjust knob (clockwise to increase vacuum and counter-clockwise to decrease vacuum). Refer to the vacuum gauge while setting the desired level of vacuum. To accurately read the gauge, block the patient end of the hose or cap off the collection bottle and allow the gauge to reach a stable vacuum reading. 
  • Position the suction tubing appropriately for patient and begin suction
  • When the liquid in the collection bottle reaches the safe full level (800 cc), the float shut-off will activate automatically and suctioning will stop.
  • Switch the ON/OFF switch to the OFF position to shut off the suction motor . 
  • Unplug the aspirator when not in use. 
  • Remove the liquid from the collection bottle after the motor stops turning completely . (Please refer to the following CLEANING INSTRUCTIONS section for directions to clean the collection bottle). 
  • Switch the ON/OFF switch to the ON position and start to run the aspirator again.
    Note: 
    - If the unit sustains a severe drop, the validity of the gauge must be checked. Inspect the float shut-off to ensure it still functions properly when the liquid in the collection bottle reaches the safe full level.
    - Always transport unit with vacuum adjust knob rotated fully clockwise in case the unit is dropped.
    Caution:
    - When automatic float shut-off is activated, empty the contents of the collection bottle immediately. Further suction could cause damage to the vacuum pump.
    - Should fluid be aspirated back into the unit, contact your Graham-Field authorized distributor for a detailed inspection, as the vacuum pump may have been damaged.


Cleaning Instructions:
Clean Collection Bottle:

 

  • Turn off the aspirator, allow vacuum to drop, then disconnect AC power source. 
  • With the collection bottle still in the holder, remove the collection bottle lid. The bottle can now be taken out of the holder to be emptied. NOTE: Empty the collection bottle immediately after each use. 
  • Clean the collection bottle after each use by using one of the following three methods:
    - Wash in a hot water/dishwashing detergent solution and rinse with clean, hot tap water . Then wash in one part vinegar to three parts hot water solution. Rinse with hot tap water and air dry . 
    - Wash with rubbing alcohol and air dry.
    - Wash with a commercial (bacterial-germicidal) disinfectant; follow disinfectant manufacturer;s recommended instructions and dilution rates carefully. 

Clean Aspirator and Collection Bottle Lid:

 

 

  • Disconnect the aspirator from all external power sources. 
  • Wipe the aspirator housing and collection bottle lid with a clean cloth and any commercial (bacterial-germicidal) disinfectant.
    Caution: Do not submerge unit in water as this will result in severe damage to the vacuum pump.

Clean Tubing:

 

 

  • Disconnect the tubing from the unit. 
  • Rinse the tubing thoroughly after every use by running hot tap water through it, followed by a solution of one part vinegar to three parts hot water. 
  • Rinse with hot tap water and air dry. 
  • Keep the outer surface of the tubing clean by wiping with a clean, damp cloth. Warning: The waste in the collection bottle, which could cause respiratory infection, must be handled according to medical waste disposal regulations. The operator must wear gloves and a medical-class mask before cleaning the collection bottle and tubing or handling the waste unit and/or accessories.


Warning WARNING: Cancer and Reproductive Harm - www.P65Warnings.ca.gov

 

GF John Bunn 800 EV2 Aspirator Specification

Particulars Dimensions
Width 4.9"
Depth 15.3"
Height 10"
Weight 11.6 lbs
Power Requirements 115 VAC, 60Hz
Vaccum Range 0 - 560 mm Hg
Power Consumption 60 watts
Operating Flow 0-40 LPM
Bacteria Filter HEPA
Operating Temperature and Humidity 32ºF to 104ºF (5º to 40º C)
Operating Humidity 95% RH

Warranty

2 Year limited warranty