Particulars |
Specification |
Size |
212 mm × 168 mm × 313 mm (8.3" × 6.6" ×12.3") |
Weight |
4.66 kg with one 12 cell battery |
Power requirements |
AC adaptor: 100-240V AC(+/- 10%)
50-60 Hz in, 24V DC, 6.25A out
DC adaptor: 11.5 - 16V DC in
19V, 7.9A out |
Purity |
87% - 96% at all settings |
Maximum oxygen discharge pressure |
20.5 psi |
Inspiratory trigger sensitivity |
-0.12cm/H2 0 |
Humidity range |
5% to 93% ± 2% non-condensing |
Operation Temperature |
5°C (41°F) and 40°C (104°F) |
Storage Temperature |
-20°C (-4°F) and 60°C (140°F) |
Setting |
Adjustable in 0.5 increments from 1.0 to 6.0 in pulse mode and from 0.5 to 2.0 in continuous mode |
Noise level |
38 dB(A) tested to Prüfmethode
14-1 03/2007 MDS-Hi
42 dB(A) tested to ISO 3744 |
Alarm types |
Low oxygen purity
No breath detected
Low battery
Service required |
Battery duration |
Approx. 4 hours with a single battery or 8 hours with 2 batteries at 18 BPM |
Operating altitude |
0; to 9000; relative to sea level (0 km to 2.7 km), 1060 down to 700 mbar |
Type of protection (electrical) |
Class II |
Degree of protection (electrical) |
Type BF |
Degree of protection (water) |
IP22 in carry bag (protection against small objects and tilted dripping water)
IP20 out of carry bag (protection against small objects and no protection against water entering the concentrator) |
Degree of safety (flammable anaesthetic mixture) |
Not suitable for use in the presence of a flammable anaesthetic mixture |
Alarm sound pressure range |
65 to 85 dB(A) |
Alarm system delays |
Less than 10 seconds after detection (low oxygen alarms if oxygen is less than 82% volume fraction at specified environmental conditions) |
Oxygen concentrator status indicator |
High priority alarm that indicates when oxygen concentration drops below 82% |