Precision Medical EasyDial Reg Oxygen Regulator

Precision Medical EasyDial Reg Oxygen Regulator

Brand/Manufacturer: PRECISION MEDICAL, INC
Hpfy Stores Save
  • FSA Approved
  • Made in USA

Precision Medical EasyDial Reg Oxygen Regulator is intended to regulate flow of oxygen from a high pressure oxygen cylinder. It has a built-in pressure relief valve, as well as a protected contents gauge. The units easily accessible flow adjustment knob is large and knurled making flow adjustment a cinch, even for the physically challenged. Its unique design minimizes normal flow variation that occurs as tank pressure drops.

Precision Medical Oxygen Regulator Highlights

  • Safe and accurate 
  • Compact and Lightweight 
  • Flexible and versatile

Precision Medical Regulator Features

  • Accurate: Unique design minimizes normal flow variation that occurs as tank pressure drops
  • Compact and lightweight
  • Ergomonically Correct Plastic T Handle
  • Color coded gauge protector
  • Easygauge reg comes in two body designs and two flow ranges, they are flexible and versatile
  • Low force Viton yoke seal provides a better tank-to-reg‚ seal
  • Lightweight aluminum body and cap
  • Contents gauge on the front and the flow gauge on the top
  • CGA connection: Yoke (870) style
  • 4-1/4" long
  • Weight: 7.8oz

EasyFial Regulator User Manual

What to buy with EasyDial Oxygen Regulator?


How to Operate EasyDial Oxygen Regulator?

  • Check cylinder valve to be sure it is not bent or damaged. A damaged valve could leak or fail, or it may not make a tight connection when cylinder is placed into use. 
  • Make sure the Regulator and cylinder valve are free of oil, grease or other contaminants.
  • Before installing Regulator, crack cylinder valve by opening the valve slightly and closing to remove any foreign particles.
  • Be sure Regulator is in the OFF position. Align OFF with the Indicating Pointer.

A. For CGA 870 connection (Yoke):
  • Be sure the high pressure inlet port has a Yoke Seal. 
  • Attach Regulator to the oxygen cylinder post valve and align index pins with mating holes.
  • Hand tighten T-handle. 

B. For CGA 540 connection (Nut and Nipple):
  • Attach the Regulator to the oxygen cylinder valve and tighten with wrench. 
  • Position the Regulator so the cylinder valve opening is pointing away from the user and any other person(s). 
  • Slowly open cylinder valve counter clockwise, until completely open. 
  • Turn dial until Indicating Pointer is pointing to the prescribed flow. 
  • Continue procedure as prescribed by your healthcare professional. 
Note:
Precision Medical, Inc. strongly recommends the use of kink proof patient tubing. When the Dial Regulator is not in use, turn dial until the Indicating Pointer is pointing to OFF and close cylinder valve.
To remove Dial Regulator from cylinder:
  1. Completely close oxygen cylinder valve.
  2. Turn dial until the Indicating Pointer is pointing to OFF".
  3. Slowly loosen T-handle or Nut and remove Regulator.

How to clean Precision Medical Regulator?

  • Disconnect all connections before cleaning.
  • Clean exterior surfaces of the Dial Regulator with a cloth dampened with disinfectant with Sporicidin R or equivalent.
  • Wipe dry with a clean cloth.
  • Store the Regulator in a clean area free from grease oil and other sources of contamination.

Precaution
  • Do not use cleaning solutions.
  • Do not attempt to repair the Dial Regulator. All repairs must be performed by Precision Medical, Inc.

EasyDial Regulator Specifications

ParticularsSpecifications
Accuracy10% of indicated setting @ 70°F (21°C)
Transport and Storage Temperature Range-4°F (-20°C) to 140°F (60°C)
Inlet Pressure Range300 psi to 3000 psi (2068 to 20684 kPa)
Ignition and Fault Tolerancemeets ASTM G175-03


Flow RangePosition Increments
0 lpm to 4 lpmOff, 1/32, 1/16, 1/8, 1/4, 1/2, 3/4, 1, 1-1/2, 2, 3 and 4
0 lpm to 8 lpmOff, 0.25, 0.5, 1, 1.5, 2, 2.5, 3, 4, 5, 6 and 8
0 lpm to 15 lpmOff, 0.5, 1, 1.5, 2, 3, 4, 5, 6, 8, 10 and 15
0 lpm to 25 lpmOff, 0.25, 0.5, 1, 2, 3, 4, 6, 8, 10, 15 and 25

EasyDial Reg Warranty

5 year limited warranty

Customer Reviews

Please enter your and your friends email address below.
Fields marked with * are compulsory
 
*
*
*
*


Submit Cancel
The selected product information has been successfully emailed to your friend.

Thank you for using www.shopnebulizer.com